Description / Abstract:
This Standards Publication establishes definitions, quantitative
measurements of performance characteristics, and reporting
techniques for the specification of the following Gamma camera
parameters:
a) Intrinsic Spatial Resolution
b) Intrinsic Spatial Linearity
c) Intrinsic Energy Resolution
d) Intrinsic Flood Field Uniformity
e) Multiple Window Spatial Registration
f) Intrinsic Count Rate Performance in Air
g) Intrinsic Spatial Resolution at 75 kcps
h) Intrinsic Flood Field Uniformity at 75 kcps
i) System Spatial Resolution without Scatter
j) System Spatial Resolution with Scatter
k) System Planar Sensitivity and Collimator Penetration and
Scatter
l) Detector Shielding
m) System Count Rate Performance with Scatter
n) System Alignment
o) SPECT Reconstructed Spatial Resolution without Scatter
p) SPECT Reconstructed Spatial Resolution with Scatter
q) System Volume Sensitivity
r) Detector-Detector Sensitivity Variation
s) Whole-body System Spatial Resolution without Scatter
The following types of medical radionuclide imaging instruments
are included in this standard:
a) Single detector, single crystal planar Gamma cameras
b) Single detector, single crystal tomographic Gamma cameras
c) Multiple detector planar and tomographic Gamma cameras
d) Whole-body Gamma camera devices
e) Discrete pixel detector Gamma cameras
The following types of medical radionuclide imaging instruments
are not included in this standard:
a) Coincidence imaging Gamma cameras or systems (these are
covered by NEMA NU 2-2012, "Performance Measurements of Positron
Emission Tomographs").
b) All medical radionuclide imaging devices not included
above